Scaramuzzi, F., Gas loading of deuterium in palladium at low temperature. J. Alloys and
Compounds, 2004. 385: p. 19.
Gas loading of deuterium in palladium at low temperature
F. Scaramuzzi
INFN-LNF, Via E. Fermi 40, 00044 Frascati (Rome), Italy
Received 12 February 2004; received in revised form 20 April 2004; accepted 20 April 2004
Abstract
The experimental technique presented in this article is aimed at measuring the absorption of
hydrogen or deuterium gas in a thin palladium sample while the system is at low temperature. A
result for deuterium is described, consisting in the measurement of the equilibrium loading ratio
X (called also D/Pd ratio, atomic), as a function of pressure, on a palladium film 3.6 μm-thick at
150 K. Values of X up to 1 have been measured at pressures lower than 1 bar. The electric
resistance of the palladium sample also has been measured as a function of temperature and of X,
and the results are reported.
Keywords: Hydrogen storage materials; Metals; Thin films; Gas-solid reactions; Low
temperatures
1. Introduction
Let us first define a quantity which we will need frequently in the following: the loading ratio
X, called also H/Pd (D/Pd) ratio, which is the atomic ratio between H (D) and Pd in the
compound. Loading hydrogen or deuterium in palladium to high values of X has been pursued
for decades, initially in order to investigate the particular behavior of this system as far as
superconductivity is concerned. It is well known that palladium is not superconducting, while its
hydride (deuteride) shows a superconducting transition if X exceeds a critical value of about 0.8,
and the critical temperature increases as X increases [1]. More recently the need for high loading
ratios (≥0.9), in particular for deuterium, has been claimed as a necessary condition in order to
investigate some of the phenomena known under the name of cold fusion [2].
Obtaining loading ratios higher than 0.7 is not easy. The value of X at equilibrium is a function
of pressure and temperature. Fig. 1a, taken from ref. [3], shows a set of isotherms for palladium
deuteride, in which the value of X at equilibrium is plotted as a function of pressure P. It is
evident that X increases with increasing pressure and decreasing temperature. In order to attain
high values of X, different methods have been used, such as electrolysis (also at low
temperatures), gas loading at high pressures (up to tens of kbar) and ion implantation; they are
thoroughly reported in ref. [1].
Fig. 1b has been obtained by cutting the graphs in Fig. 1a with an isobar at 652 torr, and shows
the behavior of X as a function of temperature. The result is encouraging, in view of attaining
high loading ratios; it has been the starting point for this experiment. A preliminary note, mainly
dedicated to testing the cryogenics of this experiment, has been published earlier [4].
When working at low temperatures, the diffusion coefficient of hydrogen (deuterium) in
palladium becomes much smaller, and reaching equilibrium can require long times. This problem
has been faced by one of the features of the method presented here. With it, it has been possible
to extend the measurements as long as necessary, until thermodynamic equilibrium was reached,
even though many days were necessary for each experimental point.
Fig. 2 describes schematically the method used: a volume V0 of about 20 cm3, is connected
through a capillary to the cell C (volume of about 3 cm3), in which the sample is located. The gas
in V0 is kept at constant (almost) room temperature, while the cell can be cooled to temperatures
down to 30 K (in the present configuration); its temperature can be held constant for long times
(up to weeks, months). The experiment consists of putting a known amount of deuterium gas in
the system, and then deducing from the measurement of the pressure P the amount of absorbed
deuterium, and thus the loading ratio X.
Figure 1. The equilibrium D/Pd ratio for deuterium in palladium as a function of pressure and temperature:
(a) is taken from ref. [3] and (b) is obtained by cutting (a) with an isobar at 652 torr.
Figure 2. Scheme of the method used for the measurement of the D/Pd ratio.
Figure 3. The “boustrophedonic” shape of the palladium serpentine. Note the six pads for the electrical
connections.
2. Description of the apparatus
2.1. The sample
The sample is a thin film of palladium, deposited by IBAD (Ion Beam Assisted Deposition)1
on a 1 mm-thick glass plate of 25 mm × 35 mm, on which a very thin layer of chromium had
been previously deposited (rf-sputtering, 400 Å). Patterning of the sample is carried out using a
previously fabricated mask for optical lithography; standard chemical etching techniques are
used to obtain the pattern, which has the shape of a “boustrophedonic” serpentine (see Fig. 3).
This geometry permits the measurement of the electrical resistance of palladium as a function of
temperature and of loading ratio, for each of the five sections in which the serpentine is divided.
Here are the data of the serpentine: length 2.5 m, width 100 μm, thickness 3.6 μm, mass 12.1 mg,
electric resistance of the unloaded sample at room temperature about 1000 Ω.
2.2. The cryostat
A closed-circuit gas refrigerator, of the Cryodyne type2 has been used. It produces refrigeration
at two stages, with the following typical (indicative) values: the first stage can absorb a thermal
power of 50 W at 50 K, the second stage a power of 5 W at 20 K. With no thermal charge
applied, the second stage can reach the temperature of 10 K. It is a rugged automated device,
which can work up to 6 months without servicing.
Fig. 4 shows the scheme of the cryostat built around the Cryodyne refrigerator. An outer
container S0 allows the vacuum insulation necessary in order to reach low temperatures. A small
activated charcoal cryogenic pump is thermally attached to the second stage of the Cryodyne.
Once a preliminary vacuum has been obtained by a rotary pump, the connection with the pump is
closed and the vacuum is insured by the charcoal adsorption; a pressure of 10-6 mbar can be
maintained for months. Inside S0 a first shield S1 (aluminum) is thermally attached to the first
stage of the Cryodyne: the scope is to drastically reduce the heat inlet from the outer can S0,
which is mainly by radiation. A second shield S2 (copper) is connected with the second stage of
the Cryodyne through a calibrated thermal link R1 (a copper wire of appropriate size). With the
help of a thermometer, of a heater, and of a commercial thermoregulator3 it is possible to keep
this shield at constant temperature at any value between 300 and about 20 K. The cell, made out
1
The sample was fabricated by Sandro Barbanera, whom the author takes the occasion for thanking, at the “Institute
of Photonics and Nanotechnologies” of the Consiglio Nazionale delle Ricerche (CNR) at Rome.
2
CTI-Cryogenics, Model 1020C Cryodyne Cryocooler.
3
Lake Shore Cryotronics Inc., Model 340 Temperature Controller.
of copper, is positioned inside S2 and is connected to it by another thermal link, R2. Also on the
cell there are a thermometer, a heater and a thermoregulator.
Figure 4. The scheme of the cryostat.
Figure 5. The cell thermoregulation for a period of about 1 day.
The idea is to keep the cell at a temperature slightly higher than that of S2. The presence of the
thermoregulated S2 shield insures a more efficient thermoregulation of the cell. A graph of the
cell temperature as a function of time for a time duration of roughly 1 day is shown in Fig. 5.
This system permits to keep the cell at any temperature between room temperature and about
30 K for long periods. A simple modification (the addition of a third shield) would permit to
attain temperatures close to 10 K, and the use of a more modern refrigerator would permit
temperatures close to 4 K.
The cell is made of two massive copper flanges, joined with the help of an indium vacuum
gasket. The glass plate on which palladium is deposited is placed in its center; this allows
dismantling the cell without touching the rest of the gas circuit. The six electrical connections for
the resistance measurements are realized through ceramic-to-metal feedthroughs with the help of
indium gaskets placed between the feedthrough and the palladium pad (see Fig. 3).
2.3. The gas circuit and the measuring protocol
Fig. 6 shows in detail the gas circuit, and helps understanding its logic. The thermoregulated
box contains the two valves (pneumatically operated) vi and ve and the head of the pressure
meter.4 The volume V0 between the two valves has been measured. The capillary starts at vi,
enters the Cryodyne vacuum tank (S0) and leads to the cell.
Figure 6. The gas circuit.
The volume of the capillary's two sections (from vi to S0 to the cell) was measured during the
set-up. The volume of the cell was measured as well, and this measurement was repeated every
time the cell was disassembled.
The experiment described hereafter consisted in the measurement of the equilibrium D/Pd ratio
with the sample at 150 K, for three different values of the pressure. The procedure was the
following. Once the system is evacuated, valve vi is closed, and through valve ve deuterium gas
is admitted to the system at the wanted pressure; then, ve is closed. At this point, all the gas
admitted to the system is contained in V0. Since volume, pressure and temperature are known, it
is possible to calculate the amount of gas, say the number of moles n, contained in the system.
Once the cell is stabilized at the wanted temperature (in this case at 150 K), vi is opened. The gas
will expand reaching the cell and will be partially absorbed by the palladium sample. It is
possible to monitor the absorption by the change of pressure. Once the pressure has reached a
value constant in time, we assume that the absorption is complete and that the system is in
equilibrium. At this point vi is closed. Taking note of the amount of gas present in V0, which can
be deduced from the residual pressure, more gas is introduced through ve up to the second value
chosen for the pressure. It is now possible to calculate the amount of gas added to the system.
The sum of this quantity and that introduced for the first step gives the total amount of gas now
present in the system. Then the same procedure described before is repeated, reaching a second
value of the equilibrium pressure. In this experiment three points have been taken, corresponding
to three values of n, and thus of P. These values of P will permit the calculation of the D/Pd
ratio, as explained in the following.
4
MKS Instruments Inc., measuring head MKS Baratron, Model 622A, 1000 torr full scale.
3. Description of the measuring method
3.1. Thermodynamics
The principle of the measurement presented here is simple, but its realization requires a certain
attention. We have stated that the number of moles of gas present in the system is known. Once
vi is open, if there is no absorption, the gas will distribute itself in the circuit so as to satisfy its
equation of state. If we know the distribution of temperature in the various volumes constituting
the circuit, we can write this equation, and thus we are able to calculate the value that pressure P
will attain at equilibrium. If there is absorption, the pressure will reach a lower value. The
difference between these two pressures, the one without absorption (calculated) and the one with
absorption (measured), will permit the calculation of the amount of gas that has been absorbed.
This, in turn, provided that we know the mass of palladium, will allow calculating the loading
ratio X.
Let us analyze the geometry of the gas circuit (see Fig. 6) from the point of view of the
temperature distribution. The content of the thermoregulated box, i.e., volume V0 plus the section
of the capillary running inside the box, are at a known constant temperature T0. The section of
capillary running from the box to S0 is exposed to room temperature; this temperature (T1) is
(slowly) varying, and is measured and monitored. The section of capillary running from S0 to the
cell is exposed to various temperatures. Its volume is about 0.4 cm3 over a total volume of the
circuit of about 23 cm3. It is thus reasonable to use the following approximate solution. We
consider two points along the capillary's path, of which we measure and monitor the temperature;
these are the first stage of the Cryodyne and the shield S2. We create thermal contacts between
these points and the capillary (see Fig. 6). In so doing we have divided this part of the capillary
in three subsections. We make the hypothesis that we can consider these subsections as three
volumes at constant (average) temperature, the temperature being the arithmetic average between
the extreme temperatures of each subsection. With this hypothesis, and remembering that the cell
is at constant known temperature, the whole system is divided in six sections, two at constant
temperature (box and cell), one at slowly varying but known temperature, and three small ones
for which we assume an average temperature; for each section we know volume and
temperature. Let us now write the state equation for each of these sections; in a first
approximation we will use the ideal gas equation, and we will justify it later (see Appendix A).
We will attribute to the six volumes and to the corresponding temperatures the subscripts from 0
to 5. Describing an equilibrium condition, we can write the following equations:
n
i
 n ni 
i
PVi
RTi
i  0,, 5
It is a system of seven equations in the seven unknowns n0, ..., n5, P. The equation on the left is
the expression of a mass conservation principle: the sum of the masses contained in each volume,
expressed as number of moles, must be equal to the total mass present in the system, n, which we
know. What we are interested in is the value of P. If we substitute the six equations on the right
in the one on the left, we obtain:
P
nR
i Vi / Ti 
i  0,, 5
(1)
A test has been performed in order to check the validity of the approximation here described, and,
in general, the validity of the method. A “blank” experiment has been performed, in which there
was no palladium. It consisted of introducing deuterium gas in the system, and measuring its
amount (which was n = 6.86 × 10-4 mol). Then the temperature of the cell was held constant at
various values, as long as necessary, so that thermodynamic equilibrium could be reached. This
was done at the temperatures of 300, 250, 200, 150, 102 and 45 K while cooling the cell, and was
repeated at the same values while returning to room temperature. The whole experiment lasted 1
week. The results of this measurement are summarized in Table 1.
The first two rows show the values of pressure measured in the two series of measurements,
going down (Pm↓) and up (Pm↑) in temperature. In the third row the average between them is
calculated (Pm). In the fourth row is reported the calculated value of the pressure, Pc. (In the
calculation the effect of the thermal expansion of the system, mainly of the cell, was taken into
account.) In the last two rows the measured and the calculated values are compared. We note
that:
1.
2.
The difference between the values measured in the two series is less than 0.3%.
The difference between measured and calculated values is less than 1%.
This result is quite satisfactory. The first statement shows that the measurements are well
reproducible. The second has to be looked at in the light of the intrinsic errors connected with
such a measurement. In fact, in order to calculate P, absolute values of T and V must be used.
These, as well as the measurement of P, are obtained by instruments characterized by an absolute
error. The sum of the errors referring to the three quoted quantities, as indicated by the
constructors, amounts to about 1%. Thus, the result is well compatible with the characteristics of
the instruments. We can conclude that the method proposed for the calculation of the pressure is
thus validated.
Table 1
Check of the proposed calculation method: comparison between measured
(Pm) and calculated (Pc) values of the pressure P; pressures are in mbar
Pm↓
Pm↑
Pm
Pc
ΔP = Pc - Pm
ΔP/P (%)
T(K)
300
754.6
752.5
753.55
758.0
4.45
0.59
250
731.0
729.0
730.0
735.3
5.3
0.73
200
699.2
698.5
698.85
704.0
5.15
0.74
150
652.3
652.0
652.15
656.5
4.35
0.67
102
581.6
581.7
581.65
583.6
1.95
0.34
45
411.0
411.0
411.1
0.1
0.02
3.2. Data acquisition
A Lab View system has been used, in order to permit recording the following quantities as a
function of time:







The temperature of the thermoregulated box, T0.
The ambient temperature, T1.
The temperature of the first stage of the Cryodyne, Tc1.
The temperature of the S2 shield, TS2.
The cell temperature, Tc = T5.
The pressure in the system, P.
The resistance of the five sections of the palladium serpentine: R1, R2, R3, R4, R5 (the last
section, R5, was broken a little time after being exposed to deuterium, and was shortcircuited).
The normal acquisition frequency was 1 reading/min, but occasionally also other values were
used, from 1 reading/s to 1 reading/5 min.
4. The experiment
4.1. Measurement of the loading ratio
The main feature of the experiment consisted in the measurement of the deuterium absorption,
i.e., the atomic D/Pd ratio, established in equilibrium in the palladium sample, while it was kept
at the temperature of 150 K. The cell was held at this temperature without interruption for the
whole duration of the experiment, which was about 1 month, with the temperature constant
within ±0.01 K. The amount of deuterium absorbed, expressed as a change Δn of the number of
moles of gas, n, present in the system, was calculated starting from the difference ΔP between
the calculated pressure and its measured value. In order to do so, formula (1) of Section 3.1 was
used. The D/Pd ratio was measured in equilibrium for three values of the pressure P,
corresponding to three values of n, the number of moles present in the system. Hereafter these
measurements are briefly described:



First point: n1 = 1.34 × 10-4 mol. Once the palladium sample was exposed to the gas, it
was necessary to wait for 9 days in order to reach equilibrium, i.e., a constant value of the
pressure; its final value was P1 = 89.9 mbar. Fig. 7 shows the variation of the pressure in
this time interval. The value of the loading ratio resulted to be D/Pd = 0.743.
Second point: n2 = 4.76 × 10-4 mol (this is the total mass of gas present in the system).
After admitting more gas into the cell, 8 more days were necessary in order to reach
equilibrium. The final value of the pressure was P = 418.4 mbar. The value of the loading
ratio resulted to be D/Pd = 0.885.
Third point: n2 = 9.43 × 10-4 mol. Also in this case it was necessary to wait for 8 more
days in order to reach equilibrium. The final value of the pressure was P = 869.0 mbar.
The value of the loading ratio resulted to be D/Pd = 1.023.
These results are summarized in Fig. 8, in a graph of the D/Pd ratio as a function of the
pressure P. The error bars will be explained below (see Appendix B).
Figure 7. The pressure P as a function of time in the 9 days relative to the first point.
Figure 8. The result of this experiment in a graph showing the D/Pd ratio as a function of the pressure P at
150 K.
Figure 9. The electrical resistance of section R1 as a function of temperature T, in the absence of absorbed
deuterium.
4.2. Measurement of the electrical resistance
Before admitting deuterium gas into the cell, a measurement of the electrical resistance of the
sample as a function of temperature was performed in the interval 300–30 K. The measurement
has been made in a dynamic mode, while the temperature was changing, though very slowly, in a
first ramp down from 300 to 30 K, and then in a ramp up from 30 to 150 K. The whole
measurement lasted 25 h. The result for the first section R1 is shown in Fig. 9 (the other sections
behave in a similar way). Note that in the same graph both series of data are reported, those in
the ramp down and those in the ramp up, and it is not possible to perceive a difference between
the two. This shows that the choice of a dynamical mode did not influence appreciably the
results.
In Fig. 10a the resistance of the same section R1 is reported as a function of time during the
first loading episode. It is not possible to correlate the value of the resistance with that of the
D/Pd ratio, but for the final equilibrium condition. It is however suggestive to note that towards
the end of the episode the value of the resistance reaches a maximum, and then starts to drop;
this is better shown in Fig. 10b, representing an enlarged section of Fig. 10a. It is well known [5]
that at room temperature the resistance of a PdDX sample reaches a maximum for X ≈ 0.75, and
this is the value that the loading ratio was approaching in this experiment towards the end of the
first episode.
Figure 10. (a) The variation of the resistance R1, as a function of time, during the 9 days of the first episode of
loading, (b) An enlarged section of the previous graph, showing clearly the maximum in the curve.
The ratio between the maximum value of the resistance, Rmax, and the value of the resistance in
the absence of absorbed deuterium, R0, is Rmax/R0 = 221.7/84.1 = 2.64: this value is, as expected,
rather higher than the value measured at room temperature, which is about 2.
The values of the resistance of the five sections in which the sample is divided are summarized
in Table 2.
The first two columns report the values of the resistance of the unloaded sample, i.e., the
sample without deuterium, at 300 and at 150 K. The following three columns report the value of
the resistance of the loaded sample, at 150 K, at equilibrium, for the three points described
above. The last column gives the maximum value attained by the loaded sample (just before the
end of the first point).
Fig. 11 shows a graph of the resistance R1 as a function of the D/Pd ratio. The curve shown
was constructed on the few experimental points at hand.
Table 2
The electrical resistance (in ohms) of the sections in which the palladium sample is divided (see
Fig. 3), as a function of temperature and D/Pd ratio: the first two columns on the left refer to the
unloaded sample, at 300 and 150 K; the following four columns refer to the loaded sample at
150 K, with reference to the three points described above, plus the maximum value attained, Rmax
R1
R2
R3
R4
R5
Unloaded sample
300 K 150 K
159.7
84.1
211.8
110.8
207.2
108.0
238.0
124.6
199.1
107.7
Loaded sample at 150 K
First point
Second point Third point Rmax
220.5
217.2
214.3
221.73
292.9
289.1
285.3
294.67
286.4
282.9
279.2
289.0
332.3
327.9
323.5
334.2
-
Figure 11. The resistance R1 as a function of the D/Pd ratio.
5. Future perspectives and conclusions
The measurements reported in this article suggest a few considerations and indicate potential
trends for the technique introduced here.
a. There are two main reasons for the choice of the temperature of 150 K for the experiment
reported here. For one thing, it was possible to use the equation of state for ideal gases,
without worrying about corrections, as shown in Appendix A. In addition, lower
temperatures could present longer absorption times, due to the sharp decrease of the
coefficient of diffusion. A guideline was the result presented in ref. [6], which shows that
absorption is practically absent at temperatures lower than 150 K, even though the data
presented there are taken in different conditions, i.e., at high pressures and in a dynamical
mode, while the temperature is changing at a rate of 2 K/min. The choice of 150 K was
considered a good compromise for this measurement. Nevertheless, it would be
interesting to investigate the behavior of such a system at lower temperatures.
b. One of the most interesting features of the method described in this paper is the
possibility of performing measurements under constant conditions for quite long times.
This has made it possible to assure that thermodynamic equilibrium had been attained by
the system, and that the distribution of deuterium in palladium was rather uniform, an
important condition in order to give a meaningful value of the D/Pd ratio. A comparison
with previous data can be performed by looking at Fig. 1b. The isobar from [3] shown in
this graph is purposely taken at the pressure of 652 torr, corresponding to 869.0 mbar, the
pressure of the third point presented in this paper. It can be noted that in the graph the
value expected for X at the temperature of 150 K is X ≈ 0.83, much lower than the value
X ≈ 1.02 found in this experiment. It is reasonable to cast doubts on the uniformity of the
deuterium distribution in the samples of ref. [3]; a higher concentration of deuterium in
the outer layers of the sample would justify such an error. This is supported by the fact
that the sample used in that measurement was 300 μm-thick, much more than the 3.6 μm
of the sample used in this experiment. This in turn suggests to check with this method the
critical temperature Tc for the superconductive transition of PdHX and PdDX; the use of a
more modern refrigerator, able to reach the temperatures of liquid 4He, would permit
performing this check in an easy way.
c. It is interesting that it has been possible to attain, at the temperature of 150 K, the value 1
for the D/Pd ratio at a pressure lower than 1 bar. It is reasonable to hope that it could be
possible to obtain higher values at higher pressures: this would permit investigating the
presence of a third phase, the γ-phase, of the palladium-deuterium (hydrogen) compound,
which has been described in the literature [7].
d. The behavior of the electrical resistance as a function of the D/Pd ratio is at the same time
comforting and slightly surprising. On the one hand, there is a confirmation of the value
of X corresponding to the maximum in the curve of Fig. 11. It is worthwhile to quote the
paper by Geerken and Griessen [8], where it is stated that for all temperatures below 300
K the maximum takes place at X ≈ 0.75, and graphs are shown with this feature at 300, 20
and 4.2 K. On the other hand, the further decrease of the resistance beyond the maximum
is much smaller in this experiment than what is reported at room temperature.
e. The curve shown in Fig. 7, representing the change in pressure of the system in the
presence of absorption, and in particular its final “exponential” section, could be useful to
get information about the coefficient of diffusion of deuterium in palladium, as a function
of temperature, of pressure and of X, provided that a reasonable model is taken into
consideration.
Acknowledgements
The author is grateful to the Fusion Division of the ENEA Frascati Research Center for
permitting the experiment described in this paper to be carried out in its laboratories. In
particular, the author is deeply grateful to Luciano Mori, Fabrizio Marini, Luigi Verdini and
Miriam Veschetti for the help provided in the construction of the apparatus and in the
management of the experiment. The author expresses also his gratitude to Antonella De Ninno,
Antonio Frattolillo and Paolo Tripodi for useful discussions.
Appendix A. Non-ideality of the gas
The equation of state of ideal gases does not take into account the interactions between atoms
(or molecules), except collisions. A well known equation of state that tries to correct the ideal
gas equation, by evaluating the effect of interactions, can be written as follows:
PV = n(RT + BP + CP2 + ∙ ∙ ∙).
where B, C, and the following coefficients take the name of “virial”. The second virial
coefficient, B, is relevant at low temperatures; the third, C, is relevant at high pressures. Thus, for
this experiment, we can restrict ourselves to consider B.
The dependence of B on temperature for deuterium is reported in Fig. 12 [9]. It is evident that
B is positive, quite small and almost constant between 400 and 200 K, then it crosses the zero at
about 120 K, and becomes negative with a much higher slope at lower temperatures. This
feature, with the hope of not needing to make corrections to the equation of state, contributed to
the choice of 150 K for this first experiment.
Figure 12. The second virial coefficient, B, for deuterium, as a function of temperature T.
Figure 13. Comparison between values of the pressure in the system measured in the same experimental
conditions at two different times. The first measurement (a) was taken at the end of the experiment. The
second (b) was started 16 days later and extended for 3 more days. In the meanwhile the system had been the
object of other tests at different temperatures (lower than 150 K), and eventually brought again to 150 K.
Table A.1
Comparison between the values of the pressure calculated following the equation of state of ideal
gases, Pig, and those obtained by taking into consideration the second virial coefficient, Pvir
T(K) Pig (bar) Pvir (bar) ΔP (bar) ΔP/P (%)
300 0.8315 0.8319 4×10-4
0.05
30 0.3326 0.3298 2.8×10-3
0.85
However, in order to render the above considerations quantitative, a simple calculation was
performed. Considering a situation very similar to the one used in the experiment (for volumes
and geometry), and given a certain amount of gas inserted in the system, the equilibrium pressure
was calculated both with the equation of state for ideal gases, Pig, and with the virial equation
above written, stopped at the term in B, Pvir for two temperatures, 300 and 30 K. By comparing
the two sets of results, reported in Table A.1, it can be noted that the correction is negligible at
300 K (0.05%), while at 30 K it starts to be relevant, with an error of 0.85%.
It is reasonable to conclude that for this experiment the use of the equation of state of ideal
gases is a correct choice.
Appendix B. Sources of error
In this section we will list a series of considerations intended to help the evaluation of the
possible errors in the measurements presented here, in particular those relative to the D/Pd ratio.
B.1. Instrumental errors
In Section 3.1 we have shown that the absolute errors due to the instruments used for the
measurement of pressure, volume and temperature result in an error lower than 1% in the
calculation of the pressure P. This error is transmitted in the determination of the D/Pd ratio,
becoming increasingly higher for each of the three loading episodes. In any case, it has been
evaluated that it does not exceed 4% in the last loading.
B.2. The sample
The error in the determination of the mass of palladium is on the order of 1%, and it has to be
added to the instrumental errors.
B.3. Consequences of a leak in the system
This is a real risk. Imagine that the system has a leak, for example, towards the Cryodyne
vacuum system. In time this would cause a decrease of the pressure in the system, thus altering
the whole measurement. Besides the many satisfactory tests performed in order to check that the
system was leak-tight, a further check is reported here. Fig. 13a shows the measurement of the
pressure P in the system as a function of time at the end of the experiment. In Fig. 13b the same
measurement is taken after 16 days and extended for more than 3 days (after performing a certain
number of checks on the system at lower temperatures, and eventually bringing the cell again up
to 150 K). The comparison between the two measurements shows a difference of less than 2
mbar on a value of 868 mbar, with an error of about 0.2% in 18 days. It is thus reasonable to
exclude this feature as a possible cause of error.
The error bars shown in Fig. 8 have been evaluated in the light of what has been said in this
section.
References
[1] B. Stritzker, H. Wiihl, Superconductivity in Metal-Hydrogen Systems, a chapter in
Hydrogen in Metals II (Application-Oriented Properties), Springer-Verlag, 1978, p. 243.
[2] M.C.H. McKubre, S. Crouch-Baker, A.M. Riley, S.I. Smedley, F.L. Tanzella, in:
Proceedings of the 3rd International Conference on Cold Fusion, Nagoya, Japan, 21-25
October 1992, Universal Academic Press, Frontiers Science Series no. 4, 1993, p. 5.
[3] E. Wicke, G.H. Nemst, Ber. Bunsenges. Phys. Chem. 68 (1964) 224.
[4] E. Del Giudice, A. De Ninno, A. Frattolillo, G. Preparata, F. Scaramuzzi, P. Tripodi, in:
Proceedings of the 8th International Conference on Cold Fusion, Lerici (SP), Italy, 21-26
May 2000, Editrice Compositori, Bologna, Conference Proceedings no. 70 of the Italian
Physical Society, 2000, p. 211.
[5] G. Bambakidis, R.J. Smith, D.A. Otterson, Phys. Rev. 177 (1969) 1044.
[6] H. Hemmes, A. Driessen, R. Griessen, M. Gupta, Phys. Rev. B 39 (1989)4110.
[7] S.A. Semiletov, R.V. Baranova, Yu.P. Khodyrev, R.M. Imamov, Sov. Phys. Crystallogr. 25
(6) (1980) 665.
[8] B.M. Geerken, R. Griessen, J. Phys. F: Met Phys. 13 (1983) 963.
[9] P. Clark Souers, Hydrogen Properties for Fusion Energy, University of California Press,
1986, p. 55. The author takes the occasion for thanking G.T. (Terry) McConville, Physics
Department University of Cincinnati, for pointing out this text and for useful discussions.